Advancements in Measuring Under-Screen Camera Display Applications Using ø2.1 mm Small Spot Probe

 

Today, many smartphones use front-facing cameras as part of the full-screen display, positioning the camera behind the screen. This configuration requires two different pixel configurations to exist on the same display, making display gamma correction more complicated. This new positioning can cause a discrepancy in the brightness and color between the area in which the camera sits behind the screen and the other display area in question. Konica Minolta Sensing addresses this challenge with its Display Color Analyzer CA-410 Small Spot Probe. When used in conjunction with a fixture assisting accurate measurement, it can measure the ø2.1 mm spot area directly in front of the camera and compare it to the adjacent area approximately 10 mm away at the same time.

USC-illust-800x300

Challenges with Under-Screen Camera Applications

The challenge of the under-screen camera (USC) display application exists with the uniformity issue between the screen located above the front-facing camera area and the rest of the screen. Pixel configurations are different between the two locations, and this will lead to a discrepancy between the two locations on the same screen for luminance and chromaticity. To correct this discrepancy, you need to perform the gamma correction for the USC area separately from the rest of the screen on the same smartphone.

To keep the USC area and the rest of the screen seamless with no color discrepancy, you need a gamma correction for the USC area to allow the color to match the adjacent area. Keeping a consistent color is more complex considering that the OLED displays have color non-uniformity in general, depending on the screen areas. If the display is not uniform between the USC area and the center screen area of the smartphone. In that case, the color difference will make the boundary between the two areas conspicuous, leading to a less than ideal appearance.

What are the Challenges in Creating this Uniformity Between these Two Areas of the Display?

To measure the screen area above the USC is the first challenge; this requires an instrument that can measure a minimal area. Secondly, finding the correct center point is key in measuring the USC area. Taking a measurement off-center can be skew the results. Finally, simultaneous measurements and distance between these measurements should be as close as possible. Simultaneous measurement with dual probes saves operation time, one of the most critical factors considered on the production line. Seamless displays require instruments that can measure and accommodate this new technology. Konica Minolta Sensing offers instruments that can assist in this application.

Konica Minolta’s CA-VP402 Small Spot Probe Addresses the Need of:

  • Measuring extremely small areas with its ø2.1 mm spot measurements

  • Provides a wide luminance measurement range covering 0.002 up to 6000 cd/m2 (or nits) with high accuracy and repeatability performance for xy chromaticity

  • Initiates fast gamma measurement (see video for demonstration)

  • Capable of FMA /JEITA/ VESA flicker measurements with a wide frequency measurement range of 0.25 up to 200 Hz

  • The ability to display waveform measurements through flicker measurement results in a waveform graph and power spectrum graph with detailed measurement information, including maximum flicker value (dB), detected frequency (Hz), and luminance max/min/average values. This feature helps test and realize the low-consumption performance of the PWM-driven display, as shown in OLEDs. (see video for demo)


New Developments

In addition, Konica Minolta Sensing can also discuss the concept of our USC display application measurement system with the use of two
CA-VP402 probes integrated into a fixture, which enables the measurement of the USC area and its adjacent area side by side at the same time.
(See exhibit A)

 

   Screen Shot 2022-12-08 at 1.04.20 PM

Exhibit A. Concept of USC Display Application Measurement System
(Application example only - fixture is not for sale rather used for illustrative purposes only)

 

This fixture has a unique optical system applied to realize the simultaneous measurement for both areas at a 10 mm distance. The optional software for this system helps with positioning the probe at the right place on the center point of the area where the camera sits underneath the display by finding each peak luminance of the X (horizontal)-axis and the Y (vertical)-axis. (See exhibit B)

 

revised fixture image with new textLeast_Square Method Graph

Exhibit B. Concept of USC Display Application Measurement System
(Application example only - fixture is not for sale rather used for illustrative purposes only)

 

Please feel free to browse our complete portfolio of light and display instruments.

Should you have any questions, please contact us.

 